Specifications
UM-20006520DE, Rev. A
Page 70
System Specifications
Output parameters
Treatment laser beam
Laser type
Holmium:YAG (Ho:YAG), pulsed
Laser wavelength
2.1 µm
Laser Classifications
FDA classification
Class IV medical device
EN 60825 classification
Class 4 laser
Power
Lumenis Pulse 50H:
1.0 to 50W
Lumenis Pulse 100H:
1.0 to 100W
Energy
0.2 - 3.5 Joules
Repetition rate
Lumenis Pulse 50H:
5 to 25 Hz
Lumenis Pulse 100H:
5 to 53 Hz
Pulse width
0.15 to 0.6 ms
Aiming beam
Laser Type
Green DPSS, continuous wave
Laser wavelength
532 nm
Laser Classifications
FDA classification
Class IIIa laser
EN 60825 classification
Class 3R laser
Power
5 mW maximum
Input power recommended service
Voltage
200/208/230 VAC ±10%
Frequency
50/60 Hz
Current
• 30A @ 208/230 VAC @ 60 Hz
• 30A @ 200/208/230 VAC @ 50 Hz
Utility connection
Single-phase grounded outlet
Cooling
Method
Internal water-to-air heat exchanger
Cooling air Requirements
Minimum 46 cm (18 in) from walls
Physical parameters
Size (W × D × H) cm (in.)
46 × 91 × 99 cm
(18 × 36 × 39 in.)
Weight
200 kg (441 lbs.)
Power cable length
7 m (23 ft.)
Footswitch cable length
5.2 m (17 ft.)
Environmental requirements (operating conditions)
Temperature
10 – 30°C / 50– 86°F
Humidity
30 to 75% at 30°C/86°F