Routine Periodic Maintenance
UM-20006520DE, Rev. A
Page 57
Routine Periodic Maintenance
Regular cleaning, inspection, testing, and repair are the basis of any
effective preventive maintenance program. Such a program helps keep the
system in top working order and ensures the reliability of safety interlocks
and fail-safe mechanisms.
A recommended routine inspection and maintenance schedule is provided
below.
Inspection/Service
Frequency
Performed By
Remarks
Routine exterior
cleaning.
As required by hospital/
clinic protocol.
Hospital/Clinic
Staff
None
Inspect cables and all
external surfaces for
damage.
Weekly
Hospital/Clinic
Staff
If damage is found, call Lumenis
Customer Service.
Inspect electrical
connections.
Weekly
Hospital/Clinic
Staff
If damage is found, call Lumenis
Customer Service.
Check remote interlock
connection and
emergency stop button.
Weekly
Hospital/Clinic
Staff
If interlock and/or button do not
perform as required, call
Lumenis Customer Service.
Inspect/replace the
debris shield
Weekly or if required due
to low output energy.
Hospital/Clinic
Staff
If output energy is still low after
replacing the shield, call
Lumenis Customer Service.
Deionizer and particle
filters replacement.
Annually
Lumenis
Service
Must be performed only by
Lumenis-authorized technical
personnel.
Electrical safety checks.
Annually (or as required
by institutional
procedures).
Lumenis
Service
Must be performed only by
Lumenis-authorized technical
personnel.
Check and perform
energy detectors
calibration procedure.
Annually, or as required if
system does not perform
to specifications, or
occurrence of error
messages.
Lumenis
Service
Must be performed only by
Lumenis-authorized technical
personnel.