4.7. SETUP PAGE 2
35
• Tens Place (0-6, 6 only available if minutes value is 9)
• Hundreds Place (0-9)
Auto Shutdown Time (Default 0 - off ) -
A short press of MODE to adjust the value. Available
values are:
• Ones Place (0-9) hours
Backlight Threshold -
A short press of MODE to adjust the value. Available values are:
• LO - Backlight comes on at a higher ambient light level.
There are two different options for Backlight Threshold Low, which can be set through
the software. Locate the parameters in the Dev tab. Enter the desired number into the
"Value" field.
–
Low Light Turn On - This value will set the threshold that the backlight will turn
on at. (This should be lower than the Low Light Turn Off value.)
–
Low Light Turn Off - This value will set the threshold that the backlight will turn
off at. (This should be higher than the Low Light Turn On value).
• HI - Backlight comes on at a lower ambient light level.
There are two different options for Backlight Threshold High, which can be set through
the software. Locate the parameters in the Dev tab. Enter the desired number into the
"Value" field.
–
High Light Turn On - This value will set the threshold that the backlight will turn
on at. (This should be lower than the High Light Turn Off value.)
–
High Light Turn Off - This value will set the threshold that the backlight will turn
off at. (This should be higher than the High Light Turn On value).
• OFF - Disables backlight
4.7 Setup Page 2
RATE/MAX Units and Minimum Display (Default 0 cpm) -
A short press of the ON/OFF to
select the value to adjust, and MODE to adjust the value. Available values are:
• Number of Decimal Places (0, 1, or 2)
• Minimum Units - See List Below
December 2021