2.8. COUNT MODE OPERATION
23
• If other operational modes are available, a short press of the MODE button will cancel
the current Count Timer and move to the next available operational mode.
• By default, the LOG button is unused during a count.
When the Count Timer has Finished:
• The display will show either the accumulated total for c, or d, or the timed ratemeter
average for cps, cpm, dpm, or Bq.
• If datalogging is enabled and logging the count results was previously requested, the
results will be logged to the log buffer.
• A short press of the
α
−
β
button will switch between the alpha, beta, and alpha+beta
displays.
• a long press of the ON/OFF button will reset the Count Timer and initialize for a new
count.
• If an alarm condition occurred during the Timed Count, a continuous audio tone will
sound as the ALARM display indicator will already be on. To acknowledge and turn
off the continuous tone alarm audio, a short press of the ON/OFF button is required.
To clear the alarm condition and reset the Count Timer, a long press of the ON/OFF
button is required.
• If other operational modes are available, a short press of the MODE button will move
to the next available operational mode.
• By default, a short press of the LOG button will log the current display.
Figure 2.10: COUNT Mode Displaying COUNT Timer of 5 Minutes, 30 Seconds
December 2021