4.5. MODEL 3002 LIST OF PARAMETERS (IN ORDER)
31
eters. If more than one parmeter can be changed on the page, short press the ON/OFF to
advance through the available parameters then used the MODE to change the parameter.
For example:
α
Rate Mode Alarm Value to 150 cpm.
Enter the Setup Menu.
• Use the MODE button to advance to P-3.
• Press the
α
−
β
button to enter the menu.
• Press the
α
−
β
button to get to the fourth of the six menu items. This is the parameter
setting for the “
α
RATE/MAX Mode Alarm Point.”
• Press the ON/OFF button to advance to the tens digit.
• Press the MODE button to increment the value to 5.
• Press the ON/OFF button to advance to the hundreds digit.
• Press the MODE button to increment the value to 1.
• Press the ON/OFF button to advance to the decimal point.
• Press the MODE button to until no decimal is displayed.
• Press the ON/OFF button to advance to the multiplier.
• Press the MODE button to until cpm is displayed.
• Press and hold the
α
−
β
button until the instrument returns to normal operation.
The list below shows the setup pages and the parameters in order, on each page.
4.5 Model 3002 List of Parameters (in order)
Page 1
• High Voltage Setting
• Overload
• Response Time
• Auto-Response Rate
• Operational Modes
• Count Time