4.3. ENTERING SETUP MODE
29
Setting
Setup Parameter
Default
Value
Notes
P4-6
Unit 2 -
β
Loss of Count
Time
60
Seconds
P5-1
Datalogging Mode
1
Mode 1
P5-2
Real Time Clock Month
1
P5-3
Real Time Clock Day
7
P5-4
Real Time Clock Year
14
2014
P5-5
Real Time Clock Hour
16
24-Hour
Mode
P5-6
Real Time Clock Minute
30
P5-7
Real Time Clock Second
00
Table 4.1: Setup Parameters
4.3 Entering Setup Mode
To enter Setup Mode, power down the Model 3002 then turn the unit back ON. When the
instrument has begun normal operation, three short presses of the MODE button will place
the instrument into Setup mode.
Entry to Setup Mode can be confirmed when the numeric portion of the display shows P-1,
indicating the first setup page is selected. If you simply wish to view the parameters, select
the desired Setup Page by a short press of the MODE button. A short press of the
α
−
β
button will advance through the parameters available on the selected Setup Page. To return
to normal operation, advance back to the Setup Page selection by pressing the
α
−
β
button
for at least a half second. Pressing the
α
−
β
for at least a half second again will exit, or holding
the
α
−
β
button for approximately 5 seconds will exit from anywhere in the menu.
ð
If the menu is not exited properly, any changes made will not be saved.
SETUP PROTECT:
By default, the Model 3002 is shipped with DIP Switch 2 in the OFF (back)
position and all available Model 3002 parameters may be modified in Setup mode. To pro-
tected the Model 3002 from changes in Setup mode, open the instrument, by removing the
six screws on the bottom of the can, and slide DIP Switch 2 (the rightmost switch) to the ON
(forward) position to activate Setup Protect Mode.
ð
The dipswitch has a plastic cover, which protects the switch during manufactur-
ing. This plastic cover is left on the dipswitch after assembly and is not removed
until the dipswitch is needed. If the plastic cover still remains on the dipswitch,
please remove the cover before changing the switch settings.
December 2021