Inhibitor
- 80 -
MTD75 Operation Guide
Inh
ibitor
Bleed Inhibiting
The facility has been provided to allow the user to inhibit any Bleed operations during and for a period
after, the Inhibitor dose. The period of the Bleed Inhibit can be set from 00:00 to 24:00 hours. When
the Bleed Inhibit is set to 00:00, the Bleed operation will be inhibited only for the duration of the
Inhibitor dose. Only available if the Bleed relay has been enabled.
Pre-Dose Bleeding
The MTD75 provides the user with the option to define a short Bleed operation before initiating the
Inhibitor Dose, this can be defined from 1 second to 99 minutes. NB. If a bleed inhibit occurs during a
pre-bleed operation the pre-bleed will be terminated and the Inhibitor dose initiated immediately.
Only available if the Bleed relay has been enabled.
Manual Dosing
To provide some flexibility in setting up and adjusting the dosing, the MTD75 is provided with a
“Manual Dosing” option. Selecting the “Manual Dose: Start” option will activate the relay for the time
set in the manual dose duration menu. If the user requires the manual dose to terminate before the
timer runs out select the “Manual Dose: Stop” menu.
Inhibitor Menu
Main Menu
From either of the front screen press the menu
button to show the main menu options and select
Dosing and Control Functions.
/
– Select Option
EXIT
– Return to Front Screen
– Enter Option
Dosing and Control Functions
Select Inhibitor.
/
– Select Option
EXIT
– Return to Main Menu
– Enter Option
Содержание MTD75
Страница 1: ...MTD75 Cooling Tower Controller Operation Guide ...
Страница 2: ......
Страница 5: ...Preface MTD75 Operation Guide 3 Declaration of Conformity ...
Страница 6: ...Preface 4 MTD75 Operation Guide ...
Страница 12: ...Introduction 10 MTD75 Operation Guide Introduction MXD75 Termination Overview ...
Страница 29: ...Installation MTD75 Operation Guide 27 Installation Horizontal Mounting Vertical Mounting ...
Страница 30: ...Installation 28 MTD75 Operation Guide Installation Blank ...
Страница 72: ...Biocide A 70 MTD75 Operation Guide Biocide A BLANK ...
Страница 74: ...Biocide B 72 MTD75 Operation Guide Biocide B BLANK ...
Страница 88: ...Inhibitor 86 MTD75 Operation Guide Inhibitor BLANK ...
Страница 98: ...pH Control 96 MTD75 Operation Guide pH Control BLANK ...
Страница 100: ...Alarm Relay 98 MTD75 Operation Guide Alarm Relay BLANK ...
Страница 108: ...Electrodeless Conductivity Setup 106 MTD75 Operation Guide ECS Setup BLANK ...
Страница 112: ...pH Input Setup 110 MTD75 Operation Guide pH Setup BLANK ...
Страница 118: ...Redox Calibration 116 MTD75 Operation Guide Redox Calibration BLANK ...
Страница 140: ...pH Calibration 138 MTD75 Operation Guide pH Calibration BLANK ...
Страница 142: ...Calibration Reset 140 MTD75 Operation Guide Calibration Reset BLANK ...
Страница 148: ...Current Outputs 146 MTD75 Operation Guide Current Outputs BLANK ...
Страница 152: ...Digital Inputs 150 MTD75 Operation Guide Digital Inputs BLANK ...
Страница 158: ...Configuration 156 MTD75 Operation Guide Configuration BLANK ...
Страница 160: ...Update Software 158 MTD75 Operation Guide Update Software BLANK ...
Страница 164: ...Save Restore Reset 162 MTD75 Operation Guide Save Restore Reset BLANK ...
Страница 166: ...Error Messages 164 MTD75 Operation Guide Error Messages BLANK ...
Страница 172: ...Modbus RS485 170 MTD75 Operation Guide Modbus RS485 BLANK ...
Страница 184: ...Data Logging 182 MTD75 Operation Guide Data Logging BLANK ...
Страница 246: ...Appendix F 244 MTD75 Operation Guide Appendix F Blank ...
Страница 254: ...Appendix G 252 MTD75 Operation Guide Appendix G BLANK ...
Страница 260: ...Notes 258 MTD75 Operation Guide ...
Страница 261: ...Notes MTD75 Operation Guide 259 ...
Страница 265: ......