![LTH Electronics MTD75 Скачать руководство пользователя страница 247](http://html1.mh-extra.com/html/lth-electronics/mtd75/mtd75_operation-manual_1945079247.webp)
Appendix G
MTD75 Operation Guide
- 245 -
Appe
ndix G
Appendix G – Error Messages
Internal Error Messages
E001
UNIT
Processor RAM Read/Write Error
Try switching the unit off and then on again. If the message persists, consult with
your supplier, as this unit may require to be returned for repair.
E002
UNIT
External RAM Read/Write Error
Try switching the unit off and then on again. If the message persists, consult with
your supplier, as this unit may require to be returned for repair.
E003
UNIT
Internal Setup Checksum Error
The instrument configuration has for some reason become corrupted. Try
switching the unit off and then on again. If the message persists, Reset Whole unit
from the Reset Unit Settings option in the Save/Restore menu or consult with your
supplier, as this unit may require a repair.
E004
UNIT
Output Card Setup Checksum Error
Try switching the unit off and then on again. If the message persists, consult with
your supplier, as this unit may require to be returned for repair.
E005
UNIT
Internal Outputs Setup Checksum Error
Try switching the unit off and then on again. If the message persists, consult with
your supplier, as this unit may require to be returned for repair.
E006 UNIT
Power
Fail
The Instrument has had its power turned off. Check instrument time.
E007
UNIT
Unit Setup Checksum Error
The instrument configuration has for some reason become corrupted. Try
switching the unit off and then on again. If the message persists Reset the whole
unit from the Load Default Settings option in the Save/Restore menu or consult
with your supplier, as this unit may require a repair.
E008
UNIT
Unit Store A Checksum Error
The data in Unit Store A has been corrupted. Save the current setup back to Unit
Store A in the Save/Restore menu.
E009
UNIT
Unit Store B Checksum Error
The data in Unit Store B has been corrupted. Save the current setup back to Unit
Store B in the Save/Restore menu.
E010
to
E013
UNIT Maths
Error
There has been an internal maths calculation failure. As such, it should not appear
if the software is functioning properly. The error message should clear after
approx. 5 seconds. If the error continues to be displayed consult with your
supplier, as this unit may require to be returned for repair.
E014
UNIT
Contrast Chip Error
The Contrast Chip is not operating correctly. Try switching the unit off and then on
again. If the message persists, consult with your supplier, as this unit may require
to be returned for repair.
E015
UNIT
Unit SD Card Checksum Error
The SD Card store from which the entire unit was restored from has become
corrupted. Check the unit's settings and then save the settings again to the SD
card store.
Содержание MTD75
Страница 1: ...MTD75 Cooling Tower Controller Operation Guide ...
Страница 2: ......
Страница 5: ...Preface MTD75 Operation Guide 3 Declaration of Conformity ...
Страница 6: ...Preface 4 MTD75 Operation Guide ...
Страница 12: ...Introduction 10 MTD75 Operation Guide Introduction MXD75 Termination Overview ...
Страница 29: ...Installation MTD75 Operation Guide 27 Installation Horizontal Mounting Vertical Mounting ...
Страница 30: ...Installation 28 MTD75 Operation Guide Installation Blank ...
Страница 72: ...Biocide A 70 MTD75 Operation Guide Biocide A BLANK ...
Страница 74: ...Biocide B 72 MTD75 Operation Guide Biocide B BLANK ...
Страница 88: ...Inhibitor 86 MTD75 Operation Guide Inhibitor BLANK ...
Страница 98: ...pH Control 96 MTD75 Operation Guide pH Control BLANK ...
Страница 100: ...Alarm Relay 98 MTD75 Operation Guide Alarm Relay BLANK ...
Страница 108: ...Electrodeless Conductivity Setup 106 MTD75 Operation Guide ECS Setup BLANK ...
Страница 112: ...pH Input Setup 110 MTD75 Operation Guide pH Setup BLANK ...
Страница 118: ...Redox Calibration 116 MTD75 Operation Guide Redox Calibration BLANK ...
Страница 140: ...pH Calibration 138 MTD75 Operation Guide pH Calibration BLANK ...
Страница 142: ...Calibration Reset 140 MTD75 Operation Guide Calibration Reset BLANK ...
Страница 148: ...Current Outputs 146 MTD75 Operation Guide Current Outputs BLANK ...
Страница 152: ...Digital Inputs 150 MTD75 Operation Guide Digital Inputs BLANK ...
Страница 158: ...Configuration 156 MTD75 Operation Guide Configuration BLANK ...
Страница 160: ...Update Software 158 MTD75 Operation Guide Update Software BLANK ...
Страница 164: ...Save Restore Reset 162 MTD75 Operation Guide Save Restore Reset BLANK ...
Страница 166: ...Error Messages 164 MTD75 Operation Guide Error Messages BLANK ...
Страница 172: ...Modbus RS485 170 MTD75 Operation Guide Modbus RS485 BLANK ...
Страница 184: ...Data Logging 182 MTD75 Operation Guide Data Logging BLANK ...
Страница 246: ...Appendix F 244 MTD75 Operation Guide Appendix F Blank ...
Страница 254: ...Appendix G 252 MTD75 Operation Guide Appendix G BLANK ...
Страница 260: ...Notes 258 MTD75 Operation Guide ...
Страница 261: ...Notes MTD75 Operation Guide 259 ...
Страница 265: ......