Appendix G
MTD75 Operation Guide
- 247 -
Appe
ndix G
Input Card Errors Continued
E138
pH
Sensor User Slope At Limit
The last Sensor Slope Calibration was out of limits, check sensor condition and
connections and repeat calibration. If the message persists please consult with
your supplier.
E139
pH
Sensor User Slope < Spec
The last Sensor Slope Calibration was less than the recommended specification,
check sensor condition and connections and repeat calibration. If the message
persists please consult with your supplier.
E140
pH
Sensor User Slope > Spec
The last Sensor Slope Calibration was greater than the recommended
specification, check sensor condition and connections and repeat calibration. If
the message persists please consult with your supplier.
E041
Redox
Sensor Input Over Range
E091 ECS
The sensor reading is greater than the specified upper limit, check input card
settings, Sensor condition and connections. If the message persists please consult
with your supplier.
E141 pH
E042
Redox
Sensor Input Under Range
E142 pH
The sensor reading is less than the specified limit, check input card settings,
Sensor condition and connections. If the message persists please consult with
your supplier.
E093
ECS
Temp Sensor Fault
E143 pH
The temperature sensor is reading open or closed circuit, due in most cases to a
damaged sensing element or incorrect wiring. Check that the temperature sensor
is set to the correct type in the channel setup menu. Under this condition, the unit
will default to the fixed temperature setting for compensation purposes. If the
message persists please consult with your supplier.
E094
ECS
Temp Input Over Range
E144 pH
The temperature sensor reading is greater than the specified upper limit, check
channel settings, Sensor condition and connections. If the message persists please
consult with your supplier.
E095
ECS
Temp Input Under Range
E145 pH
The temperature reading is less than the specified limit, check channel settings,
Sensor condition and connections. If the message persists please consult with
your supplier.
E096
ECS
Temp Comp Outside Limits
The temperature reading is less than 0.0C or greater than 150.0C, leading to an
error in compensation.
E047 Redox Calibration
Due
E097 ECS
The time since the last calibration was performed on this channel has exceeded
the time set in the calibration menu.
E147 pH
E048 Redox Planned
Service
Due
E098 ECS
The Planned Service interval for this unit has expired. Please contact LTH
Electronics at the details below:
LTH Electronics ltd, Chaul End Lane, Luton, Beds
LU4 8EZ
Tel. 0044 (0) 1582 593693, Fax 0044 (0) 1582 598036
Email [email protected]
NB. LTH overseas users should contact their LTH distributor – See www.lth.co.uk
for details.
E148 pH
Содержание MTD75
Страница 1: ...MTD75 Cooling Tower Controller Operation Guide ...
Страница 2: ......
Страница 5: ...Preface MTD75 Operation Guide 3 Declaration of Conformity ...
Страница 6: ...Preface 4 MTD75 Operation Guide ...
Страница 12: ...Introduction 10 MTD75 Operation Guide Introduction MXD75 Termination Overview ...
Страница 29: ...Installation MTD75 Operation Guide 27 Installation Horizontal Mounting Vertical Mounting ...
Страница 30: ...Installation 28 MTD75 Operation Guide Installation Blank ...
Страница 72: ...Biocide A 70 MTD75 Operation Guide Biocide A BLANK ...
Страница 74: ...Biocide B 72 MTD75 Operation Guide Biocide B BLANK ...
Страница 88: ...Inhibitor 86 MTD75 Operation Guide Inhibitor BLANK ...
Страница 98: ...pH Control 96 MTD75 Operation Guide pH Control BLANK ...
Страница 100: ...Alarm Relay 98 MTD75 Operation Guide Alarm Relay BLANK ...
Страница 108: ...Electrodeless Conductivity Setup 106 MTD75 Operation Guide ECS Setup BLANK ...
Страница 112: ...pH Input Setup 110 MTD75 Operation Guide pH Setup BLANK ...
Страница 118: ...Redox Calibration 116 MTD75 Operation Guide Redox Calibration BLANK ...
Страница 140: ...pH Calibration 138 MTD75 Operation Guide pH Calibration BLANK ...
Страница 142: ...Calibration Reset 140 MTD75 Operation Guide Calibration Reset BLANK ...
Страница 148: ...Current Outputs 146 MTD75 Operation Guide Current Outputs BLANK ...
Страница 152: ...Digital Inputs 150 MTD75 Operation Guide Digital Inputs BLANK ...
Страница 158: ...Configuration 156 MTD75 Operation Guide Configuration BLANK ...
Страница 160: ...Update Software 158 MTD75 Operation Guide Update Software BLANK ...
Страница 164: ...Save Restore Reset 162 MTD75 Operation Guide Save Restore Reset BLANK ...
Страница 166: ...Error Messages 164 MTD75 Operation Guide Error Messages BLANK ...
Страница 172: ...Modbus RS485 170 MTD75 Operation Guide Modbus RS485 BLANK ...
Страница 184: ...Data Logging 182 MTD75 Operation Guide Data Logging BLANK ...
Страница 246: ...Appendix F 244 MTD75 Operation Guide Appendix F Blank ...
Страница 254: ...Appendix G 252 MTD75 Operation Guide Appendix G BLANK ...
Страница 260: ...Notes 258 MTD75 Operation Guide ...
Страница 261: ...Notes MTD75 Operation Guide 259 ...
Страница 265: ......