![LTH Electronics MTD75 Скачать руководство пользователя страница 126](http://html1.mh-extra.com/html/lth-electronics/mtd75/mtd75_operation-manual_1945079126.webp)
Electrodeless Conductivity
Calibration
- 124 -
MTD75 Operation Guide
ECS Calibration
Insert Blue Loop
Attach the Blue (500
Ω
) loop resistor to the sensor as
shown, then press the enter button. The screen will
then indicate that the unit is sampling the sensor.
If the calibration has been completed successfully
then the instrument will automatically prompt for the
next loop resistor.
If the fail message appears then there has been a
calibration problem check the loop resistor, the
sensor and the cable. If all appears correct press the
“prev” button then the enter button to restart the
calibration.
PREV
– Exit Calibration Without Saving
SKIP
– Skip to Next Calibration Point
EXIT
– Exit Calibration Without Saving
– Initiate Calibration
ECS20 Series
Sensor
ECS40 Series
Sensor
Loop Resistor Path
Zero Calibration
Remove the loop resistor, and ensure that the sensor
head is located in free air, then press the enter
button. Note this calibration can take a few minutes.
When the calibration has completed successfully
press the enter button to save the calibration and
return to the channel’s main calibration menu.
PREV
– Go to Previous Calibration Point
SKIP
– Exit Calibration Without Saving
EXIT
– Exit Calibration Without Saving
– Initiate Calibration
Содержание MTD75
Страница 1: ...MTD75 Cooling Tower Controller Operation Guide ...
Страница 2: ......
Страница 5: ...Preface MTD75 Operation Guide 3 Declaration of Conformity ...
Страница 6: ...Preface 4 MTD75 Operation Guide ...
Страница 12: ...Introduction 10 MTD75 Operation Guide Introduction MXD75 Termination Overview ...
Страница 29: ...Installation MTD75 Operation Guide 27 Installation Horizontal Mounting Vertical Mounting ...
Страница 30: ...Installation 28 MTD75 Operation Guide Installation Blank ...
Страница 72: ...Biocide A 70 MTD75 Operation Guide Biocide A BLANK ...
Страница 74: ...Biocide B 72 MTD75 Operation Guide Biocide B BLANK ...
Страница 88: ...Inhibitor 86 MTD75 Operation Guide Inhibitor BLANK ...
Страница 98: ...pH Control 96 MTD75 Operation Guide pH Control BLANK ...
Страница 100: ...Alarm Relay 98 MTD75 Operation Guide Alarm Relay BLANK ...
Страница 108: ...Electrodeless Conductivity Setup 106 MTD75 Operation Guide ECS Setup BLANK ...
Страница 112: ...pH Input Setup 110 MTD75 Operation Guide pH Setup BLANK ...
Страница 118: ...Redox Calibration 116 MTD75 Operation Guide Redox Calibration BLANK ...
Страница 140: ...pH Calibration 138 MTD75 Operation Guide pH Calibration BLANK ...
Страница 142: ...Calibration Reset 140 MTD75 Operation Guide Calibration Reset BLANK ...
Страница 148: ...Current Outputs 146 MTD75 Operation Guide Current Outputs BLANK ...
Страница 152: ...Digital Inputs 150 MTD75 Operation Guide Digital Inputs BLANK ...
Страница 158: ...Configuration 156 MTD75 Operation Guide Configuration BLANK ...
Страница 160: ...Update Software 158 MTD75 Operation Guide Update Software BLANK ...
Страница 164: ...Save Restore Reset 162 MTD75 Operation Guide Save Restore Reset BLANK ...
Страница 166: ...Error Messages 164 MTD75 Operation Guide Error Messages BLANK ...
Страница 172: ...Modbus RS485 170 MTD75 Operation Guide Modbus RS485 BLANK ...
Страница 184: ...Data Logging 182 MTD75 Operation Guide Data Logging BLANK ...
Страница 246: ...Appendix F 244 MTD75 Operation Guide Appendix F Blank ...
Страница 254: ...Appendix G 252 MTD75 Operation Guide Appendix G BLANK ...
Страница 260: ...Notes 258 MTD75 Operation Guide ...
Страница 261: ...Notes MTD75 Operation Guide 259 ...
Страница 265: ......