Appendix G
MTD75 Operation Guide
- 251 -
Appe
ndix G
Communication Errors Continued
E448
OP Card Output Option Communication Failure
The Output Option Card is not responding. Try switching the unit off and then on
again. If the message persists, consult with your supplier, as the Output Option
card may require to be returned for repair.
E449
OP Card Output Option Communication Error
The Output Option Card is not Operating Correctly. Try switching the unit off and
then on again. If the message persists, consult with your supplier, as the Output
Option Card may require to be returned for repair.
Data Logging Errors
E450 UNIT
Setup
Checksum
Error
The Setup for the Data Logging has become corrupted. Check and correct the
Data Logging settings and turn the unit off and on again. If the message persists
please consult with your supplier.
E451
UNIT
Store A Checksum Error
The Store A Save for Data Logging has become corrupted. Check the Data Logging
settings in the Data Logging menu and then save the settings again in Store A in
the Save/Restore menu.
E452
UNIT
Store B Checksum Error
The Store B Save for Data Logging has become corrupted. Check the Data Logging
settings in the Data Logging menu and then save the settings again in Store B in
the Save/Restore menu.
E453
UNIT
SD Card Checksum Error
The SD Card store from which the Data Logging setup was restored from has
become corrupted. Check the Data Logging settings in the Data Logging menu
and then save the settings again to the SD card store.
Содержание MTD75
Страница 1: ...MTD75 Cooling Tower Controller Operation Guide ...
Страница 2: ......
Страница 5: ...Preface MTD75 Operation Guide 3 Declaration of Conformity ...
Страница 6: ...Preface 4 MTD75 Operation Guide ...
Страница 12: ...Introduction 10 MTD75 Operation Guide Introduction MXD75 Termination Overview ...
Страница 29: ...Installation MTD75 Operation Guide 27 Installation Horizontal Mounting Vertical Mounting ...
Страница 30: ...Installation 28 MTD75 Operation Guide Installation Blank ...
Страница 72: ...Biocide A 70 MTD75 Operation Guide Biocide A BLANK ...
Страница 74: ...Biocide B 72 MTD75 Operation Guide Biocide B BLANK ...
Страница 88: ...Inhibitor 86 MTD75 Operation Guide Inhibitor BLANK ...
Страница 98: ...pH Control 96 MTD75 Operation Guide pH Control BLANK ...
Страница 100: ...Alarm Relay 98 MTD75 Operation Guide Alarm Relay BLANK ...
Страница 108: ...Electrodeless Conductivity Setup 106 MTD75 Operation Guide ECS Setup BLANK ...
Страница 112: ...pH Input Setup 110 MTD75 Operation Guide pH Setup BLANK ...
Страница 118: ...Redox Calibration 116 MTD75 Operation Guide Redox Calibration BLANK ...
Страница 140: ...pH Calibration 138 MTD75 Operation Guide pH Calibration BLANK ...
Страница 142: ...Calibration Reset 140 MTD75 Operation Guide Calibration Reset BLANK ...
Страница 148: ...Current Outputs 146 MTD75 Operation Guide Current Outputs BLANK ...
Страница 152: ...Digital Inputs 150 MTD75 Operation Guide Digital Inputs BLANK ...
Страница 158: ...Configuration 156 MTD75 Operation Guide Configuration BLANK ...
Страница 160: ...Update Software 158 MTD75 Operation Guide Update Software BLANK ...
Страница 164: ...Save Restore Reset 162 MTD75 Operation Guide Save Restore Reset BLANK ...
Страница 166: ...Error Messages 164 MTD75 Operation Guide Error Messages BLANK ...
Страница 172: ...Modbus RS485 170 MTD75 Operation Guide Modbus RS485 BLANK ...
Страница 184: ...Data Logging 182 MTD75 Operation Guide Data Logging BLANK ...
Страница 246: ...Appendix F 244 MTD75 Operation Guide Appendix F Blank ...
Страница 254: ...Appendix G 252 MTD75 Operation Guide Appendix G BLANK ...
Страница 260: ...Notes 258 MTD75 Operation Guide ...
Страница 261: ...Notes MTD75 Operation Guide 259 ...
Страница 265: ......