Portable Temperature Calibraon Bath
FLUID 100 and FLUID 100-N
5
Aer any extended use at a temperature below zero:
• Bring the Set Point to about 95...100°C le ng the temperature calibrator work
ll the water inside the oil has completely evaporated.
Using the temperature calibrator at a temperature below zero for a long me
with high environmental humidity needs it takes a long me to evaporate the
whole water inside the oil (mor than one hour could be necessary).
Air humidity when the oil has a low temperature condenses on oil surface in the
shape of ice crystals.
At 95...100°C the formaon of foam means that the water inside the oil has
evaporated.