Portable Temperature Calibraon Bath
FLUID 100 and FLUID 100-N
4
WARNING!
Due to the fact that the temperature calibrator is a portable instrument, it is very
important to ensure that the socket has been earthen correctly when connecng
it to the electricity supply.
Carry out the mainentance and any repair operaon only with the equipment at
ambient temperature and disconnected from the electric power.
During the use of the temperature calibrator, the upper protecon grid may
overheat.
Do not touch the probe to calibrate when it is in the fluid reservoir.
Never change the configuraon parameters (3. Funcon Menu Level).
Do not operate the instrument in an excessively wet, oily, dusty or dirty
environment.
Do not connect any voltage to the input bushes (4), (5), (15).
Remove the cover before switching on and do not put it back unl the liquid has
reached ambient temperature and the temperature calibrator is turned off.
With the
with silicon oil 47V5 verwenden absolutely do never
FLUID 100-N
LR-
Cal
exceed the maximum temperature of +130°!
This equipment adopt the following devices to protect operaon from hazard:
• Protecon fuse (3).
• Thermal fuse to protect the container in case of over-temperature.
• Ground conductor.
Follow this instrucons:
• Never put anything on the top of the temperature calibrator.
• Never put fuel objects (liquids) near the calibrator.
• If the temeprature calibraon bath is operang at high pressure, a fume hood should be used
to remove any vapours given off by the hot bath fluid.
• Never put any type of liquids unless water-glycol (50%) or silicon oil inside the reservoir.
• Use common sense any me.
Aer any use at high temperature, bring the Set Point to room temperature le ng the
temperature calibrator go ll it‘s cold.