![LR-Cal FLUID 100 Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/lr-cal/fluid-100/fluid-100_operating-manual_1943524033.webp)
DRUCK & TEMPERATUR Leitenberger GmbH •
Bahnhofstr. 33 • D-72138 Kirchentellinsfurt • GERMANY
Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 • [email protected] •
• www.druck-temperatur.de
www.LR-Cal.net
MANUAL
Portable Temperature Calibraon Bath
FLUID 100 and FLUID 100-N
33
•
ADRESS
Communicaon adress: the value of this parameter is necessary to communicate from the PC to
many
temperature calibrators. The admi ed values are from 1 to 32 and once the value is
LR-
Cal
set by using the arrow-keys, it is necessary to confirm the choice by pressing the „E“-key.
•
S/N
Equipment serial number. It is set by the manufacturer and cannot be changed by the user.
•
Board S/N
Mainboard serial number. It is set by the manufacturer and cannot be changed by the user.
•
MAX. SET.
Maximum value to be set as the Set Point. It is set by the manufacturer and cannot be changed.
•
MIN.SET.
Minimum value to be set as the Set Point. It is set by the manufacturer and cannot be changed.
•
WAIT
Inial waing procedure (0/1):
0
= Aer switching on and aer inial self-test procedure, the calibrator immediately run to
last set point value chosen before turning off.
1
= Aer switching on and aer inial self-test procedure, the calibrator waits for input.
During waing, the 2nd. display line flashes. It is necessary to press any key in order to
move the calibrator from the waing posion and to choose the desired parameter
or value.
•
G2K-xx-xx
Internal firmware release number.
•
SENSOR TYPE
It indicates the type of the main (internal) sensor designed to adjust the temperature.
•
STAB
It indicates the swinging value of the temperature, which has been set to see on the display the
symbol of the calibrator steadiness „ “. The symbol light on when the temperature is stable for
¸
over 6 minutes.
• Cal_chnl
Recalibraon of the temperature calibrator: chooses the channel to be calibrated. It can assume
three values: INT, EXT, REF. Press the arrow-keays to select and press „E“-key to confirm.
NOTE: values „EXT“ and „REF“ are valid only for
versions of temperature calibrators.
-2I