MANUAL
Portable Temperature Calibraon Bath
FLUID 100 and FLUID 100-N
34
•
P1
First calibraon point. Press arrow-keys to set the value read with a precision reference
thermoemter and press „E“-key to accept.
•
P2
Second calibraon point. Press arrow-keys to set the value read with a precision reference
thermometer and press „E“-key to accept.
•
CAL: INT (Y/N)
This wring can have three different configuraons:
CAL: INT (Y/N)
if Cal_chnl is set to INT (see above)
CAL: EXT (Y/N)
if Cal_chnl is set to EXT (see above)
CAL: REF (Y/N)
if Cal_chnl is set to REF (see above)
Choose Y (yes) or N (no) with the arrow-keys and confirm by pressing the „E“-key.
Of course, a calibraon of measuring inputs „EXT“ and „REF“ is only possible at
versions.