Portable Temperature Calibraon Bath
FLUID 100 and FLUID 100-N
38
11. Communicaon protocol RS232/C
This chapter is needed only, if you like to communicate with the temperature calibrator with your
OWN soware. This instrucons are intended to be used by soware engineers only.
General characteriscs:
Baud Rate:
9600
Parity:
No
Data bits:
8
Stop bits:
1
The communicaon runs in half duplex way which means that the transmission and recepon of data cannot be
contemporeneously present. The regulator of the calibrator replies only aer receibing a command, it never replies
itself. The command and reply are ASCII character strings, as detailed forward. The communicaon program will be
able to convert ASCII to decimal to extract numeric values. The default adress is 1. Baud rate: 2400, 4800, 9600 and
1200 baud, the default value is 9600; the other parameters are standard and cannot be changed.
Variables
Read / Write
Type
0
Set point
read / write
float
1
Ramp ON/OFF
read / write
integer
2
Set point 2
read / write
float
3
Gradient
read / write
float
4
Resoluon
read / write
integer
5
Prop. Band
read / write
float
6
Integral me
read / write
float
7
Derivave me
read / write
float
8
Sensor input selecon
1 = internal probe
read / write
integer
2 = internal probe + EXTernal probe
3 = internal probe + REFerence probe
4 = internal probe + EXT + REF probe
9
Title
read / write
float
10
Units
0 = °C
read / write
integer
1 = °F
2 = K
13
Access key
read / write
float
14
Baud rate
read
float
15
Adress
read / write
float
16
Serial number
read
float
18
Max. set point
read
float
19
Min. set point
read
float
21
Wait ON/OFF
read
float
22
Switch on temperature
read
float
23
Switch off temperature
read
float
24
Version
read
float
25
EXTernal sensor type
read / write
integer
26
REFerence sensor type
read / write
integer
28
Stability range
read
float
29
Symbol of steadiness
read
float
100
Temperature (internal sensor)
read
float
105
Temperature (EXTernal sensor)
read
float
106
Temperature (REFerence sensor)
read
float
0 = Pt 100 (4-wire)
1 = thermocouple type N
2 = thermocouple type K
3 = thermocouple type J
4 = thermocouple type R
5 = thermocouple type S
6 = Pt 100 (3-wire)
7 = thermocouple type E