7
WARNING
Small parts can be swallowed.
WARNING
This device may only be operated by trained
personnel. User training will be provided by
Löwenstein Medical or Löwenstein Medical
authorized agents.
2.2 Precautions
Precautions are indicated by the term
CAUTION. Precautions warn the user for
special precautions to be observed in order
to ensure safe and effective use of the
prisma VENT AQUA respiratory humidifier.
CAUTION
Breathing tubes with a diameter of 15 mm
may only be used as a
heated
system with
temperature probe and heating wire adapter.
CAUTION
The use of accessories that are not approved
by Löwenstein Medical may impair the
operational reliability of the system.
CAUTION
The output power of the humidity can
be affected if the device is operated
outside of the specified range of operating
temperatures.
CAUTION
To avoid overheating, a breathing gas flow
of at least 5 l/min is constantly required in
the breathing tube system. If the breathing
gas supply is interrupted, the device must
be turned off. For applications that require
the patient's separation from the humidifier,
the device needs to be switched to the
"treatment interruption" mode.
CAUTION
The water used to fill the humidification
chamber must not be warmer than the
ambient temperature!
CAUTION
Before any maintenance, inspection or repair
work, the device must be disconnected from
the mains.
CAUTION
Every servicing of the device - including
any maintenance and testing - must be
performed by trained service technicians.
For details, please refer to the additions
to the technical specifications, and to the
maintenance and service manual.
CAUTION
Defective humidifiers may not be used.
Disconnect defective devices from the power
and notify hospital maintenance technicians
or the supplier.