22
7.2 Status display of the
current humidification
The LEDs in the central region show whether
the breathing gas has reached the set
humidification. The left and right orange
LED indicate that the humidification is less
than or greater than the set value. The
middle green LED indicates that the set
humidification is reached.
7.3 Configuration of
humidification levels
By holding the
button for three seconds,
it is possible to select between
humidification levels 1 (low), 2 (medium) and
3 (high). The Selected Humidification is
displayed via
prisma VENT AQUA
02.08.2017 | AWO
M 1:1 | all measures in mm
RAL 9006 white aluminum metallic
RAL 9006 white alluminum metallic
.
prisma VENT AQUA
02.08.2017 | AWO
M 1:1 | all measures in mm
RAL 9006 white aluminum metallic
RAL 9006 white alluminum metallic
7.4 The temperature ranges NHW, HWC and HWT
Operation mode
Level
Temperature near patient
NHW
1
approx. 23-25°C
(1)
2
approx. 25-28°C
(1)
3
approx. 28-31°C
(1)
HWC
1
approx. 30-31°C
(1)
2
approx. 31-32°C
(1)
3
approx. 32-34°C
(1)
HWT
1
31°C
(2)
2
33°C
(2)
3
36°C
(2)
(1)
these values apply at an ambient temperature of 23°C
(2)
these values apply for all permitted operating temperatures (18 to 28°C)