11. Cleaning
Before cleaning the humidifier with
accessories, it must be ensured that the
prisma VENT AQUA humidifier is switched off
and the power cord disconnected from the
mains. Furthermore, it should be ensured
that the unit has cooled down.
Do not sterilize or immerse the base unit
in liquids! Do not sterilize the temperature
probe!
Base unit, temperature probe and heating
wire distribution cable are to be cleaned
weekly or when there is a change of patient.
The base unit, temperature probe and
heating wire distribution cable can be
cleaned with a mild detergent (e.g. WILAsil).
Disinfection is performed by wiping it with
any of the following disinfectants:
{
{
Hydrogen peroxide (4%)
{
{
Isopropanol (17%)
{
{
CaviWipes
®
, METREX
®
RESEARCH
{
{
Incidin
®
Plus, Ecolab Deutschland GmbH
{
{
mikrozid
®
sensitive liquid, Schülke &
Mayr GmbH
{
{
perform
®
, Schülke & Mayr GmbH
{
{
quartamon
®
med, Schülke & Mayr GmbH
It is important to ensure that only a damp
cloth is used for cleaning! Do not allow
any liquids to penetrate the housing. The
disinfectant manufacturer's instructions
shall be strictly followed. In particular,
dilution, exposure times and change in the
composition have a major impact on the
cleaning process.
The respective user instructions must be
followed when cleaning and disinfecting the
accessories.