23
7.5 Respiratory flow areas
Operating mode / tube
diameter
Level
Respiratory flow area (l/min)
NHW /22 mm
1
5 to 50
2
5 to 40
3
5 to 30
HWT / 22 mm
HWC / 22 mm
1
5 to 60
2
5 to 60
3
5 to 40
7.6 Impact on the
performance
There are no known adverse reactions to the
performance of the prisma VENT AQUA via
impacts of electrocautery, electrosurgery,
defibrillation, X-rays, infrared rays, switching
pulses, magnetic fields and radio frequency
interference.
7.7 Muting the alarm
The audible alarm can be muted by pressing
the
button for 120 seconds. Then the
alarm sounds again. If the cause of the
alarm is not resolved within 10 minutes, the
unit switches to the "OFF" mode.
7.8 Treatment
interruption
The treatment interruption may be carried
out in normal operation, that is, if there is no
alarm activated, by pressing the button
for 3 seconds. The treatment interruption
lasts 3 minutes and can be prematurely
terminated at any time by pressing any
key. During the treatment interruption,
the heating plate and the heating wire (if
connected) can be controlled with half of the
power of normal operation.
During the treatment interruption, all 4 green
LEDs of the matrix display will flash.
7.9 Warm-up time
During the warm-up time of 30 minutes,
certain alarm messages are suppressed.