10.1 Causes and Solutions
No.
Display
Cause
Remedy
1
Temperature nearest
the patient Error
{
{
Check the installation
{
{
Check the temperature probe
{
{
Check the flow
2
Internal hardware
error
{
{
Check the installation
{
{
Shut off the device
{
{
Allow heating plate to cool before
turning on again
{
{
Contact the supplier
3
Peripheral element
Error
{
{
Connect or check the temperature
probe
{
{
Connect or check the heating wire
4
Chamber/heating plate
temperature Error
{
{
Check the installation
{
{
Check the temperature probe
{
{
Check the flow
filled circle with rays means LED is flashing
10.2 Alarm "OFF" Status
If a sounding alarm is not rectified within 10 minutes or the heating plate temperature
increases above 93°C, the forced shutdown of all heaters occurs. The device can be turned on
again only by restarting it and after cooling the heating plate again. This status is indicated by
diagonally-flashing LEDs (as shown below).