9. Representation of the
Guidelines
Display
Note
The ambient temperature is too high in relation to the selected setting.
This may result in reduced humidification.
The ambient temperature is higher than the set value of the tempera-
ture at the connection nearest the patient. The temperature cannot be
effectively controlled at the location.
There will be no risk to the patient!
Treatment interruption active. Reduction of the power of the heating
plate and the heating wire (if connected) to half. The treatment inter-
ruption period ends automatically after 3 minutes or by pressing any
button of the prisma VENT AQUA.
filled circle means LED is lit
filled circle with rays means LED is flashing
10. Alarms
When the device is switched on, there is an audible signal and a visual signal in the form of
yellow glowing LEDs on the display
, which verify the functionality of the alarm system.
All alarms are assigned to the middle priority.