Guidance and manufacturer's declaration - electromagnetic immunity to interference
The prisma VENT AQUA is intended for operation in an environment as specified below. The
customer or user of the prisma VENT AQUA should ensure that it is operated in such an
environment.
Radiation
resistance
IEC 60601 –
test level
Conformity
levels
Electromagnetic environment
- Guidelines
Conducted HF
interference
according to
IEC 61000-4-6
Radiated HF
interference
according to
IEC 61000-4-3
Radiated HF
interference in
the direct
vicinity of wire-
less communi-
cations devices
pursuant to
IEC 61000-4-3
3V
eff
150 kHz to
80 MHz
6V
eff
in ISM
and amateur
radio
frequency
bands
between
0.15MHz and
80MHz
10V/m
80 MHz to
2.7 GHz
contained in
Table 1
3V
6V
10V/m
80 MHz to
2.7 GHz
contained in
Table 1
Portable and mobile radio equipment
including cables should not be operated
within a shorter distance to the prisma
VENT AQUA than the recommended
separation distance calculated according
to that which applies to the transmission
frequency.
Recommended separation distance:
d =3.5/3
√
P
d = 3.5/3
√
P
80 MHz to 800 MHz
d = 7/3
√
P
800MHz to 2.5GHz
where P is the rated power of the
transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).
The field strength of stationary radio
transmitters should be less than the
compliance level
b
at all frequencies in
accordance with an on-site examination
a
.
In the vicinity of equipment marked with
the following symbol, interference is
possible.
.