20
21
11.1.3 Double circuit
11.2 System resistances
The total pneumatic resistance of the connected circuit and
of the connected accessories (e.g. humidifier, breathing
system filter) between the device and the patient connection
may not exceed the following value:
Circuits with a
diameter of 15 mm and 22 mm
:
Pressure drop < 3.2 cmH
2
O at a flow = 30 l/min (BTPS)
.
Circuits with a
diameter of 10 mm
(intended for volumes
delivered of < 50 ml):
Pressure drop < 3.2 cmH
2
O at a
flow = 2.5 l/min (BTPS)
.
The pressure drop values of the individual components can
be added to form a total resistance value which must not
exceed the value mentioned above.
Maximum error in pressure measurement: 0.0125 cmH
2
O
!
""
""
#
%
'
""
#
$
ARTICLE
NUMBER
ARTICLE NAME
FLOW (BTPS)
IN L/MIN
PRESSURE DROP IN
CMH2O
LMT 31382
LUISA, single circuit with valve, 180 cm, 22 mm Ø
30
0.11
LMT 31383
LUISA, single circuit with valve, 150 cm, 15 mm Ø
30
0.46
LMT 31384
LUISA, single circuit with valve, heated (i), autofill
chamber, 150 cm + 60 cm, 15 mm Ø
30
2.04
WM 271704
LUISA, leakage circuit, heated (i), autofill chamber,
passive valve, 150 cm + 60 cm, 15 mm Ø
30
2.03
WM 271705
LUISA, leakage circuit, heated (i), autofill chamber,
passive valve, 150 cm + 60 cm, 22 mm Ø
30
0.31
LMT 31577
LUISA, double circuit, 150 cm, 15 mm Ø
30
Inspiration circuit: 0.76
Inspiration circuit from
patient to device: 0.92
Exhalation circuit: 0.69
LMT 31581
LUISA, double circuit, 180 cm, 22 mm Ø
30
Inspiration circuit: 0.17
Inspiration circuit from
patient to device: 0.24
Exhalation circuit: 0.17