7 Ventilation modes
28
|
LM
T
68
54
1a
0
6/
20
21
7.7.1 Adjustable values in PSV mode
7.8 aPCV mode
In aPCV mode, fixed values can be set for the pressure levels
during inspiration and exhalation. In aPCV mode, the device
supports the patient in his or her ventilation, the patient can
trigger on inspiration. Both the patient’s spontaneous
breathing and the mechanical breaths triggered by the
device may predominate.
The graph shows the currently regulated airway pressure
over time.
PARAMETER
ADJUSTABLE VALUES
DESCRIPTION
IPAP
4 – 50 cmH
2
O(leakage circuit)
4 – 60 cmH
2
O (single or double circuit
with valve)
Set positive airway pressure on inspiration here.
PEEP
4 – 25 cmH
2
O (leakage circuit)
0 – 25 cmH
2
O (single or double circuit
with valve)
Set positive airway pressure on exhalation here.
Rate
2 – 60/min
Set background frequency (rate) per minute here.
Ti min
0.5 s – Ti max
Set the minimum duration of inspiration for spontaneous
breaths here.
Ti max
0.5 s – 4 s
Set the maximum duration of inspiration for spontaneous
breaths here.
Ti timed
Auto
0.5 s – 4 s
Set the duration of inspiration for mandatory breaths here.
On Auto setting, the Ti min and Ti max limits also apply to the
inspiration times during background ventilation.
Trigger
sensitivity on
inspiration
1 – 10
Auto
Set trigger sensitivity here:
1: Highly sensitive
10: Not very sensitive
Auto: Device optimizes trigger sensitivity.
Trigger
lockout time on
inspiration
0.2 s – 5 s
Trigger signals on inspiration are ignored during the set
period.
Target volume
On: 100 ml – 3,000 ml
IPAP max: 4 cmH
2
O to 60 cmH
2
O
Speed: 1 to 3
Speed of modification:
• Level 1: 0.5 cmH
2
O every 8 breaths
• Level 2: 1.0 cmH
2
O every 5 breaths
• Level 3: 1.5 cmH
2
Oevery breath
Pressure increase
1
2
3
4
Set the speed at which IPAP is reached here:
• Level 1: 100 cmH
2
O/s
• Level 2: 80 cmH
2
O/s
• Level 3: 50 cmH
2
O/s
• Level 4: 20 cmH
2
O/s
Pressure
reduction
(leakage
ventilation only)
1
2
3
4
Set the speed at which pressure is reduced to EPAP here:
• Level 1: -100 cmH
2
O/s
• Level 2: -80 cmH
2
O/s
• Level 3: -50 cmH
2
O/s
• Level 4: -20 cmH
2
O/s