7 Ventilation modes
LM
T 6
85
41
a 0
6/2
02
1
|
33
7.15.1 Adjustable parameters in MPVv mode
7.16 Overview of available ventilation modes in the circuits
PARAMETER
ADJUSTABLE VALUES
DESCRIPTION
Volume
100 ml – 3,000 ml
Set the volume delivered (V
t
) here.
IPAP
4 – 50 cmH
2
O (leakage circuit)
4 – 60 cmH
2
O (single or double
circuit with valve)
Set positive airway pressure on inspiration here.
Trigger
sensitivity on inspiration
1 – 10
Set trigger sensitivity here:
1: Highly sensitive
10: Not very sensitive
Trigger
lockout time on
inspiration
0.2 s – 5 s
Trigger signals on inspiration are ignored during the set
period.
Pressure increase
1
2
3
4
Set the speed at which IPAP is reached here:
• Level 1: 60 cmH
2
O/s
• Level 2: 45 cmH
2
O/s
• Level 3: 30 cmH
2
O/s
• Level 4: 15 cmH
2
O/s
MODES
CIRCUITS
Leakage circuit
Ø 22 mm, 15 mm
Single circuit with valve
Ø 22 mm, 15 mm
Double circuit
Ø 22 mm, 15 mm
CPAP
X
X
X
HFT
X
X
X
S
X
S/T
X
T
X
TTV-VAPS-AE
X
PSV
X
X
X
aPCV
X
X
X
PCV
X
X
X
aVCV
X
X
VCV
X
X
P-SIMV
X
X
X
V-SIMV
X
X
X
MPVp
X
X
X
MPVv
X
X
X