7 Ventilation modes
30
|
LM
T
68
54
1a
0
6/
20
21
The graph shows the currently regulated airway pressure
over time.
7.10.1 Adjustable parameters in aVCV mode
7.11 VCV mode
In VCV mode, the device triggers all breaths. You can set
fixed values for volume (Vt) and PEEP.
The graph shows the currently regulated airway pressure
over time.
7.11.1 Adjustable parameters in VCV mode
7.12 P-SIMV mode
In P-SIMV mode, it is possible to set both pressure support
for spontaneous breaths and a fixed value for the pressure
level on inspiration during mandatory breaths. In addition,
the lower pressure level is specified. The mandatory breaths
are triggered by the frequency setting.
Inspiration and exhalation phases are generally specified by
the patient’s spontaneous breathing, by the trigger
parameter, and by parameters F, Ti min, Ti max, and Ti.
The graph shows the currently regulated airway pressure
over time.
PARAMETER
ADJUSTABLE VALUES
DESCRIPTION
Volume
100 ml – 3,000 ml
Set the volume delivered (V
t
) here.
PEEP
0 – 25 cmH
2
O
Set positive airway pressure
on exhalation here.
Rate
2 -– 60/min
Set background frequency (rate) per minute here.
T
i
0.5 s – 4 s
Set inspiration time here.
Spontaneous breathing: Inspiration finishes no later than when
T
i
elapses.
Mandatory breath: T
i
is fixed.
Trigger on inspiration
1 – 10
Auto
Set trigger sensitivity here:
1: Highly sensitive
10: Not very sensitive
Auto: Device optimizes trigger sensitivity.
Trigger lockout time
0.2 s – 5 s
Trigger signals on inspiration are ignored during the set period.
PARAMETER
ADJUSTABLE VALUES
DESCRIPTION
Volume
100 ml – 3,000 ml
Set the volume delivered (V
t
) here.
PEEP
0 – 25 cmH
2
O
Set positive airway pressure on exhalation here.
Rate
2 – 60/min
Set ventilation frequency (rate) per minute here.
T
i
0.5 s – 4 s
Set the fixed inspiration time here.