7.8.1 Adjustable parameters in aPCV mode
7.9 PCV mode
In PCV mode, the device triggers all breaths. You can set
fixed values for the IPAP and PEEP pressure levels.
The graph shows the currently regulated airway pressure
over time.
7.9.1 Adjustable parameters in PCV mode
7.10 aVCV mode
In aVCV mode, fixed values can be set for volume (Vt) during
inspiration and for the lower pressure level during
exhalation.
In aVCV mode, the device supports the patient in his or her
ventilation, the patient can trigger on inspiration. Both the
patient’s spontaneous breathing and the mechanical breaths
triggered by the device may predominate.
PARAMETER
ADJUSTABLE VALUES
DESCRIPTION
IPAP
4 – 50 cmH
2
O (leakage circuit)
4 – 60 cmH
2
O (single or double circuit
with valve)
Set positive airway pressure on inspiration here.
PEEP
4 – 25 cmH
2
O (leakage circuit)
0 – 25 cmH
2
O (single or double circuit
with valve)
Set positive airway pressure on exhalation here.
Rate
2 – 60/min
Set ventilation frequency (rate) per minute here.
T
i
0.5 – 4 s
Set the fixed inspiration time here.
Trigger
sensitivity on
inspiration
1 – 10
Auto
Set trigger sensitivity here:
1: Highly sensitive
10: Not very sensitive
Auto: Device optimizes trigger sensitivity.
Trigger
lockout time on
inspiration
0.2 s – 5 s
Trigger signals on inspiration are ignored during the set
period.
Target volume
On: 100 ml – 3,000 ml
IPAP max: 4 cmH
2
O to 60 cmH
2
O
Speed: 1 to 3
Speed of modification:
• Level 1: 0.5 cmH
2
O every 8 breaths
• Level 2: 1.0 cmH
2
O every 5 breaths
• Level 3: 1.5 cmH
2
O every breath
Pressure increase
1
2
3
4
Set the speed at which IPAP is reached here:
• Level 1: 100 cmH
2
Os
• Level 2: 80 cmH
2
O/s
• Level 3: 50 cmH
2
O/s
• Level 4: 20 cmH
2
O/s
Pressure reduction
1
2
3
4
Set the speed at which pressure is reduced to EPAP here:
• Level 1: -100 cmH
2
O/s
• Level 2: -80 cmH
2
O/s
• Level 3: -50 cmH
2
O/s
• Level 4: -20 cmH
2
O/s
PARAMETER ADJUSTABLE VALUES
DESCRIPTION
IPAP
4 – 50 cmH
2
O (leakage circuit)
4 – 60 cmH
2
O (single or double circuit with valve)
Set positive airway pressure on inspiration here.
PEEP
4 – 25 cmH
2
O (leakage circuit)
0 – 25 cmH
2
O (single or double circuit with valve)
Set positive airway pressure on exhalation here.
Rate
2 – 60/min
Set ventilation frequency (rate) per minute here.
T
i
0.5 s – 4 s
Set the fixed inspiration time here.