6 Hygiene treatment and servicing
20
|
EN
LM
T
68
54
1a
0
6/
20
21
Cleaning coarse dust filter (gray filter)
1. Open filter compartment flap.
2. Remove gray coarse dust filter.
3. Wash coarse dust filter under running water.
4. Allow coarse dust filter to dry.
5. Replace coarse dust filter in the holder.
6. Close filter compartment flap.
Replacing fine filter (white filter)
1. Open filter compartment flap.
2. Remove gray coarse dust filter.
3. Remove and replace white fine filter.
4. Replace coarse dust filter in the holder.
5. Close filter compartment flap.
Cleaning exhalation module
1. To open the exhalation module compartment on the
rear of the device, turn the latch counterclockwise to the
symbol.
2. Remove cover.
3. Remove exhalation module.
4. Remove membrane from the exhalation module.
5. Wipe over exhalation module and membrane with
disinfectant.
Both parts can be disinfected in an autoclave at 134 °C
and 3.15 bar with a process time of 5 minutes
(maximum 50 cycles).
6. Check exhalation module for cracks and damage. If
necessary: Replace exhalation module.
7. Leave exhalation module and membrane to dry.
8. Put membrane back on exhalation module.
9. Replace exhalation module in the compartment.
10. Close exhalation module compartment.
Cleaning filter for cooling air fan
1. Open exhalation module compartment
.
2. Remove filter for cooling air fan.
Only the black translucent module is suitable for
cleaning. The black module is a disposable and must
be replaced.