Warning:
Disconnect the oxygen supply from the high-flow device during the
disinfection process.
Before connecting the ozone disinfection device, the high-flow device
must be disconnected from the mains. When the ozone disinfection de-
vice is in operation, no one should be in the same room for a long time.
→
Replacement of the accessories
In clinical use:
Replace the nasal cannula, the heated therapy tube and the autofill wa
-
ter chamber every 2 weeks or whenever you change patients. Replace
the air inlet filter every 2 months or after 1000 hours of operation.
In non-clinical use:
Replace the nasal cannula, the heated therapy tube and the autofill
water chamber every 8 weeks or whenever you change patients. The
home care water chamber can be reused for up to one year. Before
each use, however, check whether damage and / or leaks are visible
and replace the water chamber in the event of damage or after 1 year
at the latest. Replace the air inlet filter every 2 months or after 1000
hours of operation.
• If the high-flow device fails, contact Löwenstein Medical.
• To ensure long-term use of the high-flow device, the user must follow its
safety, cleaning and disinfection guidelines.
• For proper disposal of the high-flow device, please contact Löwenstein
Medical.
• The service life of the high-flow device is 10 years.
• See the label for production date information.
• For more about the guarantee: see the guarantee card.
Service and repair