23
Installation
Press the Start/Stop button for 3 seconds to
end the therapy and start or end the drying
mode. Remove the tube system and disconnect
the device from the power source.
Remove the tube system with the water cham
-
ber and insert the filter and the green cap as
shown in the figure.
Connect the red hose to the breathing gas
outlet and the ozone disinfection device to the
open connection.
Press the ozone disinfection device button to
start disinfection. The disinfection ends auto-
matically after 35 minutes.
To complete disinfection, remove the filter and
disconnect the therapy tube and the ozone
disinfection device from the high-flow device.
Do not use the device for at least one hour after
disinfection.
LM Flow disinfection instructions