11
To avoid interference, the therapy tube should be kept away from all
objects with electronic radiation and from electrical wires and conduc
-
tors.
→
Connecting the oxygen source
LM Flow (Fig.
III
a)
The LM Flow high-flow device can be con
-
nected to a low-pressure oxygen source
(e.g. oxygen concentrator). To do this, plug
the oxygen supply hose onto the oxygen
connector on the high-flow device (see
Figure IIIa).
LM Flow 100 (Fig.
III
b)
The LM Flow 100 high-flow device can be
connected to both a low-pressure oxygen
source (e.g. oxygen concentrator) and a
high-pressure oxygen source.
For connection to a low-pressure oxygen
source, please proceed as described under
LM Flow and Figure IIIa.
To connect to a high-pressure oxygen
source, screw the connecting nut of the
oxygen supply hose onto the thread of the
oxygen connection of the high-flow device
(see Fig. IIIb) and tighten the connecting nut
firmly (hand-tight).
→
How to connect to the power supply
Connect the power cord to the power connector on the back of the
high-flow device and insert the power plug into the socket. The high-
flow device screen will light up indicating that the high-flow device is
being properly supplied with power.
III a)
Oxygen connection
(Low pressure)
III b)
Oxygen connection
(Low pressure)
Oxygen connection
(High pressure)