The high-flow devices of the LM Flow series humidify the respiratory ther
-
apy gas with water vapor. If the regular cleaning, disinfection and replace
-
ment of parts are not carried out in accordance with these instructions,
there is a risk that bacteria will settle on and in the high-flow device, which
can infect the patient. The high-flow device must therefore be cleaned
thoroughly before and after each use. It must be cleaned and disinfected
when used on multiple patients.
→
Cleaning
Daily cleaning: Hoses, nasal cannula, water chamber. Connect the en
-
tire tube system with nasal cannula and activate the drying mode. Once
dry, remove the water chamber and rinse with clean water.
Warning:
The water chamber and the therapy tube must not be cleaned with
cleaning agents. Use a clean, lint-free, disposable cloth (moistened
with a neutral detergent) to wipe the connector between the auto-fill
water chamber and the high-flow device.
Also check that all impurities have been removed.
The power supply of the high-flow device must be disconnected before
cleaning. Do not place the device in liquids.
Cleaning and maintenance