28
7.5 Auto Power Off Default Setting
• Use “ ” Up, “ ” Down to select “ 1 “ or “ 0 “.
1 = Auto power On.
0 = Auto power Off.
• After Auto Power Off adjustment, push the “Enter” button, then
press “ESC” to save the data and return to the normal display.
7.6 Temp. Unit Default Setting
• Use “ ” Up, “ ” Down to select “1“ or “0“ .
1 = °F
0 = °C
• After adjusting the Temperature unit, push the “Enter” button,
then press “ESC” to save the data and return to the normal
display.
7.7 pH Manual Temp. Setting
• This procedure is only to adjust the manual temperature
compensation value for pH measurement.
• The lower display will show:
M. TEMP. SET
^, v Enter:Y
• Use “ ” Up, “ ” Down to select the desired manual Temp.
compensation value.
• Press “Enter“ once, then press “ESC“ to save the data and
return to the normal display.
7.8 Temp. Compensation Factor Setting
• This procedure is only used for the Conductivity function.
• Use “ ” Up, “ ” Down to select the Temp.
Compensation Factor (% per °C) of the measured solution.