6) Use “ ” button (Page 5, Fig. A-5), “ ” button to adjust the
upper display value to match the standard conductivity value.
7) Press the “Enter” button twice to save the calibration
data, and finish the calibration procedure.
• If only one calibration point is needed, just set the 2 mS
range (1.413 mS Cal.).
• A multi-point calibration procedure should always start
with 2 mS range (1.413 mS Cal.), then proceed to other
ranges (20 µS range, 20 mS range or 200 mS range) if
necessary.
CD CAL.
^, v ENTER: Y
23.5 °C
1.413