18
Display layout
Measurement Value
% O
2
24.2 °C
Temp. value Temp unit
Temp. unit
% O
2
or mg/L
20.9
5. DO (Dissolved Oxygen) Measuring and Calibration
Procedure
ATTENTION:
Make sure the Oxygen probe is filled with Elec-
trolyte!
To fill the Probe‘s Electrolyte, refer to chapter 5.3 “Probe
maintenance“, page 22.
The meter default settings are as follows:
• The display unit is set to % O
2
.
• The temperature unit is set to °C.
• Auto power off.
• Data logger sampling time: 2 seconds.