3.1 pH measurement (with manual temperature setting)
1) Attach the pH Electrode by installing the “Probe Plug“
(Fig. B-1) into the “pH Socket/BNC Socket“ (Page 5, Fig. A-16).
2) Power on the meter by pressing the “Power” button once.
3) Keep pressing the “Mode” button until the display of bottom
right shows “pH“ and “Manual Temp.“ indicator.
4)
Adjust the Manual Temp. value exactly to the same
temperature
as the solution. For procedure refer to chapter
7.7, page 28.
5) Remove cap and hold the pH Electrode body (Fig. B-2) and com-
pletely immerse the “Sensing Head“ (Fig. B-3) in the solution to
be measured and gently swirl the probe.
6) The upper display will show the pH value, the bottom left display
will show the Manual Temp. setting.
Fig. B
B-1
B-2
B-3
B-4