Peak and Valley Detection
(Maximum/Minimum Reading View)
The meter can store the lowest and highest reading values in memory. You can see these
values if you press the MAX/MIN front panel button, or by using remote contact closure switches.
The first press shows peak, with the upper left hand indicator bar lit (marked ‘Max.’). The sec-
ond press shows valley, with the lower left hand indicator lit (marked ‘Min.’). The display returns
to the running value after 2 or 3 seconds. To reset the memory, press the RESET button for more
than 3 seconds while peak or valley is being displayed. Peak and Valley values are not stored if
you switch the meter off.
How to use the MAX/MIN button to view Peak or Valley
1) Link terminal 7 to terminal 9
2) Set lockout switch ‘ON’
3) Press UP arrow key (MAX/MIN)for peak, valley, normal
Remote contact closure viewing of Peak/Valley
1) Connect a normally-open contact closure switch between terminals 7 and 9
2) Set the calibration lockout switch ‘ON
Tare Command
You can force the reading to zero by pressing the front panel tare switch for 2 to 3 seconds, or by
applying a remote contact closure. Following readings will be the ‘net’ value, offset by the read-
ing at the time the tare switch was operated. The tare value can either be stored in non-volatile
memory, or can be cleared at power-off. This is set in the Tare menu. The ZERO led will light to
show you that the unit has been tared.
Front Panel key Taring
1) Link terminal 7 to terminal 8
2) Set calibration lockout switch ‘ON’
3) Press ZERO key for 3 seconds to tare display. The ZERO led will light to show you that the
unit has been tared.
To reset the tare from the front panel, press the Tare button and the Reset button together. The
ZERO led will go out when tare has been cleared.
Remote contact closure taring
1) Connect a normally-open contact closure switch between terminals 7 and 8
2) Set calibration lockout switch ‘ON’
Logic Inputs and extra button functions
Warning:
Disconnect all power before
exposing the rear of the meter
Warning:
Logic inputs are NOT isolated
from input signal
Lockout notes:
Cal Lock switch must be ON to enable logic inputs.