21
Linearisation - theoretical scaling
Lin.O
Lin.O
Lin.1
Press OK to select
Press both left buttons briefly
You can now choose to have the
lineariser on or off
Choose THEORETICAL Calibration method - See the Calibration methods page.
theo
dc v
in.00
rd.00
ad.pt
quit
save
Confirms you are in
Theoretical Cal mode
Confirms DC voltage
range.
If DC current, this display
will show DC A
Set Value for input
signal at point 0.
Press OK when
done
Set Value for
display at point 0.
Press OK when
done
Press OK to select
Press OK if you want
to add another point
Press OK if you want
to quit / abort.
Press OK if you want
to finish and save.
in 10
rd 10
quit
save
Set Value for input
signal at point 10.
Press OK when
done
Set Value for
display at point 10.
Press OK when
done
Press OK to select
Press OK if you want
to quit / abort.
Press OK if you want
to finish and save.
Displayed value
In 00
rd 00
rd 01
In 01
In 02
In 03
In 04
In 05
In 06
In 07
In 08
In 09
In 10
rd 02 rd 03
rd 04
rd 05
rd 06
rd 07
rd 08
rd 09
rd 10
Input signal
You can set as many or few points
as you wish. 11 points available.
Lockout notes:
Cal Lock switch = OFF when changing. Switch Cal Lock to ON after changing