Maintenance
Module 8023
6-13
Power cabling must be rated at no less than 5A current carrying capability and
be routed into the control box by a suitable watertight gland. The power input
gland hole is on the extreme left of the control box when facing the controls.
Access to the ac input connections is as access to the incoming ac protection
fuse & reject relay protection fuses
on page 11
for
boards protected by a one piece safety mesh cover. For boards with the two
piece cover only the angled terminal cover access plate requires removal.
Figure 4 AC Power/Ground input connection units built before November 2001
Figure 4 AC Power/Ground input connection units built after November 2001
Live Neutral
JP301
TP303
The earth connection is
made to the control box
ground stud located just
below the edge of the
control board
Pin1
Live Neutral
JP301
TP303
The main incoming earth connection is made to the left
hand control box ground stud (in a group of 3) located
just below the right hand edge of the control board
Pin 1