4
1.3 Product Contents
The LUNA-II
™
product package contains the following components.
*
The LUNA-II
™
Automated Cell Counter with built-in printer (L40001) comes with one pack of LUNA-II
™
Printer Paper installed.
Upon receiving the product package, please inspect its contents to ensure that all parts have been included and that no
damage has occurred during shipping. The warranty does not cover damage that may occur during shipping and handling.
Any damage claims must be filed with the carrier. Contact your local distributor or Logos Biosystems if anything is missing.
1.4 Product Specifications
*
Processing time may vary according to cell type and concentration.
**
This is the minimum processing time for each focusing option at the specified concentration of HeLa or HL-60 cells.
Component
Quantity
LUNA-II
™
Automated Cell Counter
*
1
Power Cord (with AC adapter)
1
LUNA
™
Cell Counting Slides, 50 Slides (100 Counts)
1 box
USB Drive, 16 GB
1
Trypan Blue Stain, 0.4%
2 x 1 mL
LUNA-II
™
Automated Cell Counter Specifications
Instrument Type
Benchtop cell counter
Dimensions (W x D x H)
16 x 18 x 28 cm (6.3 x 7.0 x 11.0 in)
Weight
1.6 kg (3.5 lb)
Cell Concentration Range
5 x 10
4
- 1 x 10
7
cells/mL
Cell Diameter Range
3 - 60 µm (optimal range: 8-30 µm)
Cell Viability Range
0 - 100%
Image Resolution
5 MP
Image Type
TIFF
Processing Time
*
10
**
(manual focusing) or 15
**
(autofocusing) seconds at ~1 x 10
6
cell/mL
LUNA
™
Cell Counting Slide Specifications
Material
Polystyrene
Dimensions (W x D x H)
25 x 75 x 2.4 mm
Chamber Depth
100 µm
Chamber Volume
10 µL