73
7.3.5 “Calibration Step 3” is necessary for “Fluorescence Counting Calibration.” Fluorescence
standard beads are supplied by Logos Biosystems or local distributors. Load 10
l of fluorescence
calibration beads into one of the counting chambers and wait for about 1 min in order for cells to
completely settle down. Insert the counting slide and then press the “Start” button. Make sure to
load the original standard solution - do not dilute it.
<Calibration Step 3-1>
<Calibration Step 3-2>
Note
: Never turn off the Luna-FL™ cell counter during calibration, which could lead to a significant
technical failure.