43
<Unsettled Cells>
4.3.2 Insert the loaded slide into the slide port of the instrument, ensuring that the loaded chamber
is inserted first into the slide port. The instrument analyzes only the first inserted chamber. Gently
insert the counting slide to the end.
Note
: After inserting the slide, Luna-FL only reads the first chamber. To read the second chamber, it
must be taken out, rotated, and inserted again.
Note
: Make sure that the counting slide is not inserted upside-down. This may lead to sample
spilling and could severely damage the counter.
4.3.3 Select the
“
Fluorescence Cell Counting
”
button on the home screen.
<Home Screen>