Doc. Ref.No: Q875-0143/ENG
Issue 2
Met 30+ Operating Manual
12/01/10
5. USING THE SOFTWARE
This section of the manual describes what the software does and how to use it. It is
intended mainly for reference and is not meant to describe how the software would
typically be used, or why certain options are required.
5.1 Power
up
When the unit is switched on it will display the product name and bargraph. This is
called the normal running display.
TEST SETUP
Pressing the or keys will cycle the display through the product setup, status and
back to the bargraph. None of the information can be modified at this level.
Pressing the key will first show a screen showing the reject and product counts
(Standard), or reject counts and type of reject (Pharmaceutical only): S = Size of
signal, P = Phase of signal.
Met30+ Standard
Met30 + Pharmaceutical
Prod count:
31473
S 92
P 45.0
Rej count:
12
Rej count:
12
The next screen to be displayed shows detection parameters:
Met30+ 1f
Met30 + 3f
W=100 H= (32000)
W=100 H= (32000)
G=H P= 0.0°
G=H P= 0.0° F=H
W is the width, H the height and P the phase (
θ
) used for detection as shown on the
diagram below.
Note:
width and height refer to characteristics of the electrical signal
from the product, and have nothing to do with the physical dimensions of the
product.
The height value is shown in brackets if height-checking is disabled, i.e. the height is
not being used for detection and the product detection envelope is open-ended.
The other figure displayed is the gain (G) and there are two gain settings with code
version V4.xx, i.e. High and Low. With all other code versions the maximum Gain of
the system is 15. If the product effect is large, the gain will be reduced.
File Name: Q875-0143 ENG Iss2 Met30+ Operating.doc
Page 17 of 58