D9U001GT0-0101_09
59
15.1 Lifting Pole
It is possible to equip the bed with lifting pole. The lifting pole must be placed only to the bushing on the head end.
Fig. Bushing for lifting pole and infusion stand
Lifting pole variants:
■
Lifting pole for beds with fixed bed end
To ensure safe use of the lifting pole:
►
Never exceed the maximum load of 75 kg of lifting pole.
►
Never use the lifting pole for rehabilitation exercises.
►
To prevent the bed from tipping over, ensure the lifting pole does not project out from the bed.
►
Replace plastic handle every 4 years.
Lifting pole positions:
■
Over the backrest (working position).
■
Parallel to the head board (when not in use).
To install the lifting pole:
►
Insert lifting pole into corresponding bushings at the head end of the bed (corners).
►
Ensure the safety pin locks are in place.
A plastic grab handle with an adjustable strap is attached to the lifting pole.
NOTE
: The date of manufacture is marked on the grab handle.
NOTE:
LINET® recommends replacing the plastic grab handle every four years.
15.2 Infusion Stands
WARNING!
Risk of injury due to use of unsuitable accessories!
►
Use infusion stands exclusively for accessories listed in the instructions for use.
►
Only mount an infusion pump to the lower (wider) telescopic section of an infusion stand above the head board /
foot board.
►
Never mount an infusion pump to the upper (thiner) telescopic section of an infusion stand.
►
Ensure the infusion pump will not collide with any movable parts of the bed (especially Backrest part) or with the
patient. This must be verified during installation.
►
Do not use the infusion stand as driving/pushing device during the bed transport.
It is possible to insert infusion stands into the bushings at the head and foot ends of the bed.
►
Use exclusively infusion stands with 4 hooks for hanging IV bags or baskets for intravenous solutions.
►
Ensure that the weight-bearing capacity of the 4 hooks is not exceeded.
Capacity per hook: 5 kg
►