D9U001GT0-0101_09
46
Cardiac chair position (button 13)
The cardiac chair position is suitable for patients with cardiac arrhythmia and breathing difficulties.
Settings after pressing and holding the Cardiac chair button (13):
►
Calfrest tilts into lowest position, thighrest moves into the upright position (34°).
►
6 second after positioning calfrest and thighrest the backrest will move into an upright position (62°).
Trendelenburg position (12)
The Trendelenburg position serves as an anti-shock position. All parts of the mattress platform are flattened. Mattress platform tilts
head down.
CPR (Resuscitation) Position
The CPR position is for resuscitating the patient in an emergency.
Settings after pressing and holding the CPR button (11):
►
Mattress platform moves into a horizontal position.
NOTE:
For quick mechanical positioning, see CPR Backrest Release.
12.3 Handset
The handset is included as a standard feature. The handset is available with and without button illumination. The button illumination
of the illuminated handset is active when the bed is connected to the mains. The functions of both handsets are identical. Where
the handset is to be stored on the bed depends on the patient’s condition.
1. Thighrest Adjustment Button
2. Thighrest/Backrest Lock LED
3. Backrest Adjustment Button
4. Activating GO Button
5. Auto-Contour Button
6. Height Adjustment Button
7. Height Lock LED
8. Flashlight Button
9. Flashlight
Fig. Handset
To switch on the flashlight:
►
Press and hold flashlight button 8 and flashlight 9 on the top of the handset will light up.
Set the positions as follows:
►
Activate the keypad by pressing the GO button.
►
Press and hold the function button until the desired position is reached.
NOTE:
Depending on the patient’s condition, the nursing staff decides whether the patient is allowed to adjust the bed’s position.
If required, prevent the patient from adjusting the bed as follows:
►
Disable functions.
►
Unplug handset (if the bed is equipped with Plug and Play).
NOTE:
The handset can be plugged into another LINET® hospital bed with the Plug and Play slot.