ServiceManual Uno™200 mobile lift (3EN500402 Rev. 5)
9
www.liko.com
4.0
Troubleshooting
13
www.liko.com
Uno 200 • 7EN150105-05
Simple Troubleshooting
If you hear unusual sounds.
Contact Liko/Hill-Rom.
The lift does not work up/down.
Width adjustment does not work
in/out.
1. Check that the emergency stop button is
not pushed in (p. 7).
2. Check that the cables to the control unit
are correctly connected (p. 6).
3. Check that the charger cable is not
connected to an electric socket.
4. Check that the battery is charged (p. 10).
5.
If the lift still does not work satisfactorily,
contact Liko-Hill-Rom.
Battery charging does not work.
1. Check that the emergency stop button is
not pushed in (p. 7).
2. Check to ensure that the electric socket
is power-supplied.
3. Check that the charger cable is correctly
connected (p.10).
4.
If the lift still does not work satisfactorily,
contact Liko/Hill-Rom.
The lift stops in the elevated position.
1. Check that the emergency stop button is
not pushed in (p. 7).
2.
Use the electrical emergency lowering
to lower the patient to a stable underlying
surface (p 8).
3.
Use the mechanical emergency
lowering to lower the patient to a stable
underlying surface (page 8).
4. Check that the battery is charged (p.10).
5.
If the lift still does not work satisfactorily,
contact Liko/Hill-Rom.
.
.
.
.
.
.
.
.
.
Содержание Uno 200
Страница 2: ......
Страница 4: ......
Страница 26: ...ServiceManual Uno 200 mobile lift 3EN500402 Rev 5 26 www liko com...
Страница 38: ...www liko com www hill rom com Manufacturer Liko AB Nedre v gen 100 SE 975 92 Lule Sweden...