EMC Table
Guidance and manufacturer’s declaration – electromagnetic emissions
The mobile lift is intended for use in the electromagnetic environment specified below. The customer or the user of the
mobile lift should assure that it is used in such an environment.
”Essential performance according to the manufacturer: The hoist shall not move unintentionally while being submitted to
disturbances.”
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The mobile lift uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The mobile lift is suitable for use in all establishments inclu
-
ding domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Complies
Voltage fluctuations/ flicker emis
-
sions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The mobile lift is intended for use in the electromagnetic environment specified below. The customer or the user of the
mobile lift should assure that it is used in such an environment.
”Essential performance according to the manufacturer: The hoist shall not move unintentionally while being submitted to
disturbances.”
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 8 kV contact
+/- 15 kV air
+/- 8 kV contact
+/- 15 kV air
+/- 8 kV contact
+/- 15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient / Burst
IEC 61000-4-4
+/- 2 kV for power supply
lines
+/- 1 kV for input/output
lines
+/- 2 kV for power supply
lines
n/a. for input/output lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
+/- 1 kV differential mode
+/- 2 kV common mode
+/- 1 kV differential mode
n/a. for common mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short inter
-
ruptions and voltage va
-
riations on power supply
input lines
IEC 61000-4-11
0 % UT for 0,5 cycle, at 0, 45,
90, 135, 180, 225, 270 and
315 degrees
0 % UT for 1 cycle, at 0
degrees
70 % UT for 25 cycles at 50
Hz and 30 cycles at 60 Hz, at
0 degrees
0 % UT for 250 cycles at 50
Hz and 300 cycles at 60 Hz.
0 % UT for 0,5 cycle, at 0,
45, 90, 135, 180, 225, 270
and 315 degrees
0 % UT for 1 cycle, at 0
degrees
70 % UT for 25 cycles at 50
Hz and 30 cycles at 60 Hz,
at 0 degrees
0 % UT for 250 cycles at 50
Hz and 300 cycles at 60 Hz.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
[Equipment or System] requires
continued operation during power
mains interruptions, it is recom
-
mended that the [Equipment or Sys
-
tem] be powered from an uninter
-
ruptible power supply or battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
Complies
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com
-
mercial or hospital environment
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.