background image

21

www.hillrom.com

7EN140108 Rev. 4  •  2020

Quick Guide 

Prod. No

 

2000400

Liko Mobile lift system

 

(Fit´s "Holder for Quick Reference Guide", Prod. No. 2000100)

Recycling instructions

Waste of Electrical and Electronic 

 

Equipment (WEEE).

Metals

Lead battery (Pb) or 

Li-ION battery

Example: Transfer from wheelchair to bed

7EN190760 Rev.1

English

xxxxxxxxxxxxxxx

Liko AB 

Nedre vägen 100 

SE-975 92 Luleå , Sweden

www.Hill-Rom.com  I  www.liko.com

Quick Guide

Liko® mobile lift system 

Important!

Verify the maximum load on each component. 

The lowest rated component always state the 

maximum load to the composite lift system.

 

Push in the sling 

under itself, under 

the patient.

Important!

  

Never use worn or  

damaged components  

in a lift system.

Disconnect 

the sling.

Patient centered 

over the bed.

Lift the patient.

Turn in and lower 

the patient in the 

center of the bed. 

Place the sling

centered.

Finalize the transfer

Lifting and transferring  

Preparing the lift system 

 

Carefully remove the 

sling under the patient.

The lift system in the example has 

a Maximum load

 =

 200 kg* 

*stated by the mobile lift.

The transfer is

completed.

Charge  the 

lift after use!

Important! 

Before lifting,

always check the sling loops 

in the sling bar hooks.

Safety instructions

Important!

Plan the transfer and compose 

your lift system based on the 

patient need´s and safety.

 Read the User Manual for more information.

This Guide do 

not replace

 user manuals!

Mobile lift

 

Maxload: 200 kg

Sling bar

  

Maxload: 300 kg

Sling

Maxload: 250 kg

Example:

  

The lift system below is a composition of 

 

3 components, all with different maximum load;

Note! 

When lifting from floor 

use locked wheels and 

a pillow under the 

patient’s head.

(C)2016. Lik

o R&D AB. All Righ

ts R

eser

ved.

 Mobile lift towards: 

Chair / Wheelchair

Mobile lift towards:

 Bed

Mobile lift towards: 

Patient on the floor

Important! 

All sling bar

hooks shall carry the load 

for  a balanced and safe lift

.

Cross the leg support loops 

through each other.

Seek short and low 

lifting of patients 

during transfer!

Old batteries are to be deposited at the nearest recycling station or given to personnel authorized by Hill-Rom.

Hillrom evaluates and provides guidance to its users on the safe handling and disposal of its devices to aid in the 

prevention of injury, including, but not limited to: cuts, punctures of the skin, abrasions, and any required cleaning and 

disinfection of the medical device after use and prior to its disposal. Customers should adhere to all federal, state, regional, 

and/or local laws and regulations as it pertains to the safe disposal of medical devices and accessories.
If in doubt, the user of the device shall first contact Hillrom Technical Support for guidance on safe disposal protocols.

Содержание 2000045

Страница 1: ...mance and safety when using the lift Golvo 9000 mobile lift Instructions for Use Golvo 9000 Prod No 2000045 Golvo 9000 LowBase Prod No 2000049 7EN140108 Rev 4 ENGLISH 2020 QRG IMPORTANT Lifting and tr...

Страница 2: ...ctions 4 Definitions 5 Technical Data 5 Dimensions 6 EMC Table 7 Assembly 9 Disassembly 12 Operation 12 Charging the Batteries 18 Maximum Load 19 Recommended Lifting Accessories 19 Recycling Instructi...

Страница 3: ...line underneath the symbol indicate this product have been placed on the market after 2005 UL Recognized Component Mark for Canada and the United States EFUP Environmental Friendly Usage Period years...

Страница 4: ...uestions please consult the responsible assistive device technician or the supplier The lift may not be used in areas where flammable mixtures may occur for example in areas where flammable goods are...

Страница 5: ...mm 3 in twin wheels with brakes Turning diameter Golvo 9000 1330 mm 52 in Golvo 9000 LowBase 1330 mm 52 in Emergency lowering device Mechanical and electrical Lifting Speed no load 37 mm s 1 5 in s or...

Страница 6: ...9000 LowBase B4 Model A B B1 B2 B3 B4 C D D2 E F F1 G Lmax Lmin M O max min max min max min Golvo 9000 2090 1455 1185 870 480 480 1020 735 907 623 0 1100 105 25 940 1816 571 436 391 Golvo 9000 LowBas...

Страница 7: ...ge ESD IEC 61000 4 2 8 kV contact 15 kV air 8 kV contact 15 kV air 8 kV contact 15 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humi...

Страница 8: ...s the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequenc...

Страница 9: ...on distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the t...

Страница 10: ...ightening the two pre assembled M8 screws A B 8 Optional accessories Quick reference guide to be placed in Holder for Quick reference guide on the mast 7 Hang the hand control on the handle Motor for...

Страница 11: ...single beep is heard service interval activated Alternatively use the push buttons simultaneously for emergency lifting up and down on the control box 11 Place the charger cable on the hook provided...

Страница 12: ...e the lift using the push buttons on the handcontrol For raising and lowering Directional arrows show the direction of movement up down The lifting and base movement stops as soon as the push button i...

Страница 13: ...rning check cables and connections Warning is shown until repaired Overload Too much load is applied to the lift Service needed contact Hill Rom SHORT CIRCUIT CHECK CONNECTION kg Ibs service 6 4 3 5 2...

Страница 14: ...as follows Push down the red emergency lowering handle at the same time as loading the sling bar pull down the sling bar with your other hand This extends the strap and the sling bar is lowered Repea...

Страница 15: ...d in the parking panel the lift should not be raised since this could be dangerous and cause personnel injury or damage to the lift if the sling bar should release and swing out from the panel Install...

Страница 16: ...nd smooth and free from gravel debris ice and potholes Two caregivers are recommended for this task A gurney or a wheel chair should be in the immediate vicinity of the car when performing the lift Re...

Страница 17: ...caregiver must guide the sling bar and patient out of the car and ensure that the head of the patient is guided safely while exiting the car The other caregiver will need to raise the lift while simul...

Страница 18: ...box internal charger standard Plug the charger cable into mains 100 240 VAC see charger information 1 2 above The battery is fully charged after about 6 hours and the charger disconnects automaticall...

Страница 19: ...e lift is affected Before changing lifting accessories you should always ensure that the lift after change can fulfil the desired lifting height in order to manage the lifting situations for which the...

Страница 20: ...ormation LikoScale device for weighing a patient in combination with Golvo mobile lift LikoScale 350 Max 400 kg 880 lbs Prod No 3156228 LikoScale 350 is certified according to the European Directive N...

Страница 21: ...nsfer and compose your lift system based on the patient need s and safety Read the User Manual for more information This Guide do not replace user manuals Mobile lift Maxload 200 kg Sling bar Maxload...

Страница 22: ...tery capacity Check if the Li ION battery has been set in to sleep mode see chapter Operation 5 If the problem persists please contact Hill Rom The charger doesn t work 1 Check that the charger cables...

Страница 23: ...own the entire lift starting from the top down Do not use a cloth that is dripping wet To have access to all areas run the lift into the highest and lowest positions and extend the base width adjustme...

Страница 24: ...Golvo and ceiling lifts Alcohol Isopropyl alcohol 70 5 0 7 0 Viraguard Veridien Hand controls for all lifts Quaternary ammonium n Alkyl dimethyl benzyl ammonium chlorides 0 105 n Alkyl dimethyl ethyl...

Страница 25: ...o product Expected Life Time The product has an expected service life of 10 years when correctly handled serviced and periodically inspected in accordance with Liko s instructions Parts listed below a...

Страница 26: ...Copyright Liko R D AB 2020 ALL RIGHTS RESERVED Liko AB Nedre v gen 100 975 92 Lule Sweden 46 0 920 474700 www hillrom com Liko AB is a subsidiary of Hill Rom Holdings Inc...