MODBUS Register v1.48
248083389-1 Rev
B-3
40020
ASCII string
Model Name char[10], char [11]
40021
ASCII string
Model Name char[12], char [13]
40022
ASCII string
Model Name char[14], char [15]
40023
unsigned integer
Flow Rate. Divide by 100 to get rate in CFM. For example,
100 = 1CFM.
40024
unsigned integer
Record Count. Total number of records stored in the counter.
40025
unsigned integer
Record Index. Zero based index to data in 30xxx register
series. Must be lower than the record count (register 40024).
Set this index to expose a counter’s record in the 30xxx
registers. Set to -1 to retrieve last record stored in the counter.
40026
unsigned integer
Location number.
Particle Counters: Specifies location of Particle Counter. Must
be 1 to 200 (maps to location names associated with registers
40200 - 40999).
Manifold Controller: Specifies Manifold position. Values 1-32
for the Universal Manifold and values 1-6 for the MiniManifold
Controller moves the arm to that position on the manifold.
Value 0 moves arm to Home position.
40027
signed integer
Real Time Clock (RTC) [high]. Updates instrument’s real-time
clock. Works in conjunction with 40028. Displays date and
time, in number of seconds since midnight, 1/1/1970. Can be
generated by ANSI C/C++ time() function.
40028
signed integer
Real Time Clock [low]
40029
unsigned integer
Initial Delay [high]. Works in conjunction with 40030.
Number of seconds to wait before starting the first sample. Max
value is 359,999, which equals 99h 59m 59s.
40030
unsigned integer
Initial Delay [low]
40031
unsigned integer
Hold Time [high]. Works in conjunction with 40032. Number
of seconds to wait between sample periods. Max value is
359,999, which equals 99h 59m 59s.
40032
unsigned integer
Hold Time [low]
40033
unsigned integer
Sample Time [high]. Works in conjunction with 40034.
Number of seconds to sample. Max value is 86,399, which
equals 23h 59m 59s.
Table B-3 Sensor Settings Registers
Register Data
Type
Description
Содержание solair 3100
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions S3100 S5100 Boulder Counter Gen E Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual t vi 248083389 1 Rev...
Страница 14: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual ii 248083389 1 Rev...
Страница 19: ...Introduction 248083389 1 Rev 2 3 Figure 2 1 SOLAIR Dimensions Side Dimensions Top Dimensions...
Страница 22: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 2 6 248083389 1 Rev...
Страница 36: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 3 14 248083389 1 Rev...
Страница 60: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 4 24 248083389 1 Rev...
Страница 66: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual 5 6 248083389 1 Rev...
Страница 115: ...Operating Counter 248083389 1 Rev 6 49 Figure 6 64 SOLAIR 3100 STATUS Screen...
Страница 180: ...Lighthouse S3100 S5100 Boulder Counter Operating Manual D 2 248083389 1 Rev...
Страница 187: ...00...
Страница 188: ......